Abstract
A procedure for simultaneous determination of amlodipine (AML), hydrochlothiazide (HYD) and valsartan (VAL) in pharmaceutical products was established basing on spectrophotometric – chemometric method using full spectra combined with Simulan program. The procedure was proved with good repeatability (relative standard deviation (RSD) values fluctuated from 2,2% to 2,3% for AML, HYD and VAL n = 3) and good trueness (recovery variations were 90,2% - 94,6% for AML, 91,3% - 95,5% for HYD, and 102,1-104,5% for VAL, n = 3) in terms of AML, HYD and VAL analysis in phamarceuticalproduct Exforge HCT. The results obtained by the developed method were agreed with those by high performance liquid chromatography (HPLC).

This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.
Copyright (c) 2017 Array