Abstract
Dead time is one of several major parameters of a nuclear radiation measurement system. To estimate the dead time and its uncertainty for the TCS-172 gamma survey meter at the Training Center of the Nuclear Research Institute, we conducted 240 count rate measurements using the two-source method with the non-paralysable model and two standard radioactive sources, Cs-137 and Co-60. The results indicate that the dead time for this instrument was 11.29 µs, with an uncertainty of ±1.01 µs. These values align closely with those presented in the textbook and several articles.
References
- Aloka Co., Ltd. Gamma scintillation survey meter, Model TCS-172 instruction manual.
- IAEA-TECDOC-1363. Guidelines for radioelement mapping using gamma ray spectrometry data. IAEA; 2003.
- Tsoulfanidis N, Landsberger S. Measurement & detection of radiation. 4th ed. CRC Press; 2015.
- Akyurek T. A new dead-time determination method for gamma-ray detectors using attenuation law. Nuclear Engineering and Technology. 2021,53(12):4093-4097.
- Saha GB. Physics and radiobiology of nuclear medicine. 4th ed. Springer; 2013.
- Gupta TK. Radiation, ionization, and detection in nuclear medicine. Springer; 2013.
- Cherry SR, Sorenson JA, Phelps ME. Physics in nuclear medicine. 4th ed. Elsevier; 2012.
- McGregor DS, Shultis JK. Radiation detection: concepts, methods, and devices. CRC Press; 2020.
- Quyet DP. Determination of the dead time of a GM counter using the two-source method. International Journal of Science and Research Archive. 2024;13(01):790-795.
- Certificate of calibration, No. TT & CD/PKP/2000/1694, Board of radiation and isotope technology, Government of India deparment of atomic energy.
- Quyet PD. Select an equation to calculate the dead time for a radiation counter system. International Journal of Science and Research Archive. 2025;15(03):433-438.

This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.
Copyright (c) 2025 Array